Specifiche tecniche:

Datalogger di Umiditˆ Relativa Veriteq SP-2000


Specifications for the Veriteq SP 2000 line of precision relative humidity & temperature recorders.

Description: Self-powered humidity recorder complete with on-board humidity & temperature sensors, non-volatile memory, 10-year battery and precision clock.

Internal Relative Humidity Sensor

Type of sensor

Integrated high-accuracy capacitive polymer RH sensor

Initial Accuracy

+/-1%RH over 10%RH to 90%RH at +20¼C to +30¼C

+/-2%RH over 10%RH to 90%RH at -20¼C to +70¼C

Operating range

-35¡C to +85¡C, 0% RH to 100% RH (non-condensing)

Resolution

0.05% RH

1-year Accuracy

(*denotes Accuracy One Year after original deployment)

*+/-2% RH over 10% RH to 90% RH at +20¡C to +30¡C
 
+/-3% RH over 10% RH to 90% RH at -20¡C to +70¡C

Calibration Interval

Annual (recommended)

Internal Temperature Sensor

Initial Accuracy

+/-0.10¼C over +20¼C to +30¼C

+/-0.15¼C over -25¼C to +70¼C

Operating range

-35 to +85¡C (-31 to 185¡F)

Resolution

0.02¡C at 25¡C

1-year Accuracy

+/-0.15¡C over 20¡C to +30¡C
+/-0.25¡C over -25¡C to +70¡C

General Humidity Logger Specifications

Size

3.4 x 2.3 x 1" (85x59x26mm);
76g (2.7 oz)

Operating Range

-35 to 85¡C (-31 to 185¡F)
and 0% RH to 100% RH

Interfaces

RS-232 serial, USB, Ethernet, WiFi, PoE network interface available.

Mounting

Magnetic strips; 3M Dual Lockª fasteners

Software

Compatible with the following software packages: Veriteq Spectrum, Veriteq viewLinc (monitoring & alarming applications), Spectrum Excel Add-in, Spectrum API (for custom applications) and OPC Server to add loggers onto compatible monitoring systems.

Clock

Accuracy: +/- 1 min./month at -25¡ C to 70¡ C.

Electromagnetic Compatibility

FCC Part 15 and CE

Power Source

Internal lithium battery with life of 10 years at 1 min. sampling rate or longer.

Memory

Memory Type

Non-volatile EEROM

Data Sample Capacity

35,100 12-bit samples

Memory Modes

User-selectable: Wrap when memory full (FIFO), or Stop when memory full.

Memory Protection

Data retention > 20 years backup without power.

Sampling Rates

User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day.

Recording Span: 2 Channels Enabled

Sample Interval

Recording Span

10 seconds

2.0 days

1 minute

12.1 days

5 minutes

2.0 months

15 minutes

6 months

1 hour

2 years