Temperature recording and specifications for Veriteq 1000/1400 Precision Data Loggers


Specifications for the Veriteq 1000 and 1400 Precision Temperature Datalogger Recorders include:

General

Size

3.4 x 2.1 x 1" (85x59x26mm);
60g (2.7 oz)

Operating Range

-40 to 85¡C (-40 to 185¡F) and
0 to 100% RH

Interfaces

RS-232 serial port; ethernet; PoE

Mounting

Magnetic strips; velcro optional

PC Software

Compatible with the following software packages: Veriteq Spectrum, Veriteq viewLinc (for on-going monitoring & alarming), Veriteq OPC server software, Veriteq sGo (for Palm OS), Spectrum Excel Add-in, and Spectrum API (for custom applications).

Clock

Accuracy: +/- 1 min./month at -25¡C to +70¡C

Electromagnetic Interference

Meets FCC Part 15 for digital devices; meets CE requirements for radiated emissions, electrostatic discharge, & radiated susceptibility.

Power Source

Internal lithium battery with life of 10 years at 1 minute sampling rate.

Temperature Recorder Accuracy

Sensor

Calibrated
Measuring Range

Resolution

One year Accuracy

Internal

-25¡C to +70¡C

0.02¡C at +25¡C

+/-0.15¡C over +20¡C to +30¡C

+/-0.25¡C over -25¡C to +70¡C

ÒNÓ Range External
Probe

-25¡C to +70¡C

0.02¡C at +25¡C

+/-0.15¡C over +20¡C to +30¡C

+/-0.25¡C over -25¡C to +70¡C

ÒLÓ Range External Probe

-50¡C to +10¡C

0.02¡C at +20¡C

+/-0.25¡C over -50¡C to +10¡C

ÒVÓ Range External Probe

-90¡C to -40¡C

0.02¡C at -80¡C

+/-0.25¡C over -90¡C to -40¡C

Internal Temperature Sensor

Type of Sensor

Precision-tolerance epoxy-encapsulated NTC thermistor

Response Time

10 seconds in moving air (thermistor only)

Memory - SP-1000 model

Memory Type

Non-volatile EEROM

Data Sample Capacity

48,100 12-bit samples

Memory Modes

User-selectable: Wrap when memory full, or Stop when memory full.

Memory Protection

Data retention > 20 years backup without power.

Sampling Rates

User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day.

Recording Span

Recording span depends on sample interval selected and number of channels enabled. Chart below details typical sampling rates and length of time logger will retain data in memory before wrapping around or stopping (see Memory Modes).

 

Sample Interval

Recording Span (w/number of channels enabled)

 

1 channel

2 channels

 

10 seconds

5.5 days

2.7 days

 

1 minute

1.1 months

16.7 days

 

5 minutes

5.5 months

2.7 months

 

15 minutes

1.3 years

8.3 months

 

1 hour

5.4 years

2.7 years

 

Memory - SP-1400 model

Memory Type

Non-volatile EEROM

Data Sample Capacity

85,300 12-bit samples

Memory Modes

User-selectable: Wrap when memory full, or Stop when memory full.

Memory Protection

Data retention > 20 years backup without power.

Sampling Rates

User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day.

Recording Span

Recording span depends on sample interval selected and number of channels enabled. Chart below details typical sampling rates and length of time logger will retain data in memory before wrapping around or stopping (see Memory Modes).

 

Sample Interval

Recording Span (w/number of channels enabled)

 

1 channel

2 channels

3 channels

4 channels

 

10 seconds

9.8 days

4.9 days

3.2 days

2.4 days

 

1 minute

1.9 months

29.6 days

19.7 days

14.8 days

 

5 minutes

9.8 months

4.9 months

3.2 months

2.4 months

 

15 minutes

2.4 years

1.2 years

9.8 months

7.4 months

 

1 hour

9.7 years

4.8 years

3.2 years

2.4 years